Etigam BRIGHT-CHEQ Biological Indicator Monitoring System
BRIGHT-CHEQ Incubator Operation Manual
10
INCUBATOR INSTALLATION
Initialize Screen
Warming Up Screen
Main Menu
1. Follow the Printer Setup procedures in this section
of the manual to assemble and bring the printer
online.
2. Connect the printer to the incubator using the
supplied serial printer cable.
3. The incubator power pack (12V) has four
plug adapters for international use. Install the
appropriate plug adaptor for your environment
onto the incubator power pack.
4. Insert the small connector from the incubator
power pack into the matching receptacle in the
back of the incubator, then plug the power pack
into an outlet.
The incubator will turn on and the Initialize
screen will appear on the LCD indicating that
the incubator has started its initialization
routine. After initialization, the WarmingUp
screen will appear with a date and time in the
upper right hand corner of the display. The
status LEDs on the face of the incubator will
blink red in succession while the incubator
warms up.
5. Place the thermometer into the cell labeled T.
6. Wait for the status LEDs on the incubator to turn
o and for the MainMenu screen to appear.
7. Congure the incubator using the procedures
in the Conguration - Global Settings and
Conguration - Cell Settings sections of this
manual.
Note: The incubator is designed for continuous use and
does not have a power switch.
INSTALLATION