Etigam BRIGHT-CHEQ Biological Indicator Monitoring System
BRIGHT-CHEQ Incubator Operation Manual
16
SOUND CONFIGURATION
Main Menu
Setup Menu
Sound Menu
1. From the MainMenu, touch the SETUP key to enter
theSetupMenu.
2. From the Setup Menu, touch the SOUND key to
enter the SoundMenu.
3. To turn key-touch conrmation sound on or o,
touch the NAV toggle key until the desired setting
appears. When NAV is on, a sound will play each
time a key is touched on the screen.
4. To enable or disable the alarm that sounds when a
BI has turned yellow or has remained purple for the
entire incubation period, touch the ALARMS toggle
key until the desired setting appears.
5. To enable or disable the alarm that sounds when
a BI is removed from an incubation cell, touch
the REMOVE toggle key until the desired setting
appears.
6. Touch the BACK key to return to the SetupMenu.
7. If conguration is complete, touch the BACK key to
return to the MainMenu.
CONFIGURATION - GLOBAL SETTINGS